TOC is used in a variety of applications from management of wastewater
treatment plant influent and effluent, to drinking water supply
management and monitoring of impurities in process and natural
waters. And the range of applications in which this type of on-line
technology is used is expected to continue expanding. In addition,
because nitrogen is considered to be one of the factors which contributes
to the over-proliferation of organisms in enclosed bodies of water,
management of these public water sources and of treatment plant
influent and effluent also includes demand for monitoring of nitrogen
levels.
- High performance on-line TOC analyzer using the established
680°C catalyst-aided combustion and non-dispersive infrared
detection method
- Wide measurement range is variable from 0-5ppm
full scale to 0-1000ppm full scale. Measurements up to 20000ppm
are possible using the on-board dilution function
- Measurement
mode selectable between TC/NPOC (standard), TOC=TC-IC (optional)
or TOC=POC+NPOC
|