Welcome Back to Unit 4!

no peaks detected

 

Welcome back for another session on LCMS troubleshooting. In previous units, we provided a basic understanding of the different types of MS detectors and when you would use a certain type, typical preventive measures to prevent issues from arising in the first place and the most common causes of pressure issues. This week we will be exploring the causes and solutions when we don’t observe any peak.

It can be extremely frustrating to set a batch running to come back to no usable results. As such, it is always good practice to check the first injection to make sure that everything is working correctly.

If we observe no peaks in our results, we must explore the sample, the LC and the detector as potential causes.

 

1. MS Setting Issues

Issue Description Solution
Unsuitable MS Technique for Analyte The mass spectrometry may not be a suitable detection technique for the physicochemical properties of the analyte (e.g. non-ionisable or thermally unstable compounds).
Incorrect Probe or Capillary Position Improper probe distance from the orifice or incorrect capillary protrusion can result in inefficient ionisation or poor spray
Incorrect MS Parameter Settings Incorrect ion source type, acquisition mode or polarity can prevent analyte ionisation and detection
Schedule Acquisition Window Too Narrow If operating in scheduled MRM or similar modes, the analyte may elute outside the defined event window
Instrument Tune or Performance Drift Poor MS tuning or detector instability can lead to loss of sensitivity or no peak detection
Tuning file

 Fig. 1 Example of a tuning file

Adduct Formation or Matrix Suppression Adducts (Na+, K+, NH4+, etc) or matrix effects can suppress analyte signal.
Sample Below Detection Limits The analyte concentration may be below the MS detection limit

 

2. LC and Method Issues

Issue Description Solution
No Flow or Incorrect Flow to MS No mobile phase reaching the ion source due to disconnected tubing or open purge valve
LCMS method
 

 Fig. 2 Check the LCMS method settings to ensure all conditions are correctly set

Air in System or Pump Malfunction Air bubbles or pump malfunctioning can interrupt the solvent delivery which in turn will affect elution or detection of the peak
Blocked or Damaged Tubing Crimped or blocked tubing restricts flow and may stop analyte delivery to the detector
Incorrect Retention or Elution Compounds may elute in the void volume (too early) or remain retained on the column (not eluting)
Incorrect Flow Rate Flow rate errors may prevent analyte transfer to the MS or reduce sensitivity
Blocked Injection Needle A blocked injection needle prevents sample introduction to the system, often with increased pressure
Divert Valve or Waste Routing The analyte may be diverted to waste by a programmed time event
Waste valve

 Fig. 3 Valve program set to waste after 2.000 mins

 
 

The above examples should provide the most observed issues which might cause the peak to be unobserved. By systematically ruling out either the MS, LC or sample, focus can be directed to the mostly likely culprit which can then get you back up and running again once fixed.

In the next course unit, we will look into changes in peak shape and their most common causes. 

Your Shimadzu LCMS team 

 

Related Resources