Changes in Peak Shape (Part 2)

LCMS Troubleshooting Course

Part 2: Causes of Peak Shape Changes in LCMS 

This topic is divided into two parts:

  1. 1. Injection & data acquisition effects on peak shape.
  2. 2. Causes of peak shape changes (fronting, tailing, broadening) and MS-specific factors.

In the last course unit, we began to look at some of the causes of peak shape changes, which can impact on the accuracy and reliability of our data.This week, we shall look at specific peak shape issues and their causes.
 

A: Peak Fronting

Fronting

 

This is characterised by a steep peak rise and shallow tail, known as a "shark fin"

 

 
Issue Description Solution
Column Damage The column is a consumable with a lifespan. The column could just be aging, or potentially the lifespan could be reduced if working outside the manufacturer's specifications (i.e. pH or temperature range)
Sample Overloading Overloading can be caused by excess analyte saturating the column or overwhelming the ionisation capacity, thereby distorting the peaks.
Sample Solvent Incompatibility or Poor Solubility The sample diluent is integral to the peak shape. If the sample solvent is too strong, it can cause the analyte band to spread and distort.
MS Specific Considerations If the sample is too concentrated, the MS signal can max out where the detector is saturation, causing flat topped peaks. 

B: Peak Tailing

Tailing

This is observed as the opposite of a fronted peak, where there is a slow descent of the peak which is typically observed for basic species. If this behaviour is observed in neutral species, it would indicate issues with the column. 

      
Issues Description Solution
Secondary Interactions Secondary interactions between a charged analyte and a charged moiety on the stationary phase, such as residual silanol groups, can cause the peak to present a broad tail. 
Inadequate Buffering Ionisable species require adequate buffering in order to maintain suitable peak shape. Be aware that 0.1% formic acid only equates to ~2 mM which in some circumstances is not sufficient. 
Column Damage If the column is damaged or the integrity of the packed bed is impacted, the shape of all peaks could be impacted. 
Adsorption Adsorption of the analyte in tubing or fittings can cause tailing. Another common problem is chelation of the analyte with metals in the wettable flow path. 
Sample Loading As previously discussed, the sample concentration can cause an increase in peak tailing, which is due to the adsorption Langmuir Isotherm. 
Interfering Peak If there are two peaks which possess the same transitions, such as regioisomeric species, it might be possible that the two species are not adequately resolved, thereby causing a peak to appear tailed. 
Column Voiding or Degradation If the integrity of the column has 
Dead Volume If the column has not been installed correctly, it can create additional volume at the head of the column, which can cause peak issues. 
MS Specific Causes Poor ionisation or source contamination can worsen tailing by stretching ion arrival times.

C: Peak Broadening

 

Broadening

Peak naturally spread throughout the chromatographic system, which is why they produce their characteristic peak shape. However, we want these peaks to be as narrow as possible, which means we need to optimise the system to minimise system volume and dispersion. It is also important to notice if peak shapes change over the course of several analyses because this can impact on your resolution.

       
Issue Description Solution
Dead Volume An incorrectly installed column could cause band broadening, presenting as a broader peak
Detector Response Time Too High The detector settings could be set that it is artificially increasing the width of the peak.
Column Aging or Temperature Gradients text text text text text text text text text
Longitudinal Diffusion in Isocratic Runs Peak broadening due to residency time in the column is just a characteristic of isocratic runs. The longer the analyte is on the column, the broader it will be. 
Sample Loading See previous comments. 
Oven Setting Issues If the wrong settings have been used on the oven compared to the original method, peaks might change peak width. 
Mobile Phase The wrong mobile phase will give a wrong response which could include changes in peak shape as well as changes in retention time. 
Slipped Ferrule A slipped ferrule could result in poor connections which causes peak dispersion.
LC Setup The LC setup should be recorded to ensure that the same dwell volume and system volume is used for all analyses. The dwell volume is the tubing between where the two solvents meet, typically at the mixer in high pressure mixing systems or the proportionating valve in low pressure mixing systems, to the head of the column. The system volume is the volume from the point of injection through to the head of the column. Changes in these values can change peak width 
MS Specific Causes Ion suppression and spray instability can cause changes in peak width. Erratic nebuliser gas or probe positioning widens the peak response times, therefore notes should be taken to ensure the probe is positioned correctly and the spray should be investigated for stability. 

 

 

D: Peak Splitting

Splitting

This is characterised by two peak apexes instead of a single peak for one compound.

 
Issue Description Solution
Soiled Guard or Column Inlet This is commonly from insufficient sample preparation.
Sample Diluent Incompatibility with Mobile Phase The sample diluent can be too strong, and particularly for early eluters, can negatively impact peak shape, often with split peaks.
Analyte Properties The properties of the compound could cause peak splitting, such as isomers or analyte interconversion
Detector Saturation The peak shape could be negatively impacted by the detector being saturated with samples, which should be investigated and rectified. 

 

Additonal LCMS Troubleshooting for Peak Shape Issues

If peaks are missing, split, or distorted despite proper LC conditions:

Daily LCMS Peak Shape Checklist

 

In this second course unit of the changes in peak shape, the possible causes for various effects are explained. The most common cases people see in practice are tailing and peak broadening. It is imperative to improve the peak shapes we observe in order to maximise on resolution between peaks of interest and ensure impurities are not hiding underneath a peak.

In the next course unit, we will be looking at ghost peaks and contamination.

 

Your Shimadzu LCMS Team

Related Resources